8 Trouble-shooting
TopClean S CPC 30
100
EHauser
E067
M
A
I
N
T
E
N
A
N
C
E
R
E
Q
U
I
R
E
D
Set point exceeded
controller / LS 1
Dosing devices defective;
Chemical supply empty;
Measured value incorrect -> check for plausibility and
function;
Incorrect control direction set;
incorrect contact assigned;
Incorrect control function assigned
yes
no
no
E068
Set point exceeded
controller / LS 2
yes
no
no
E069
Set point exceeded
controller / LS 3
yes
no
no
E070
Set point exceeded
controller / LS 4
yes
no
no
E071
Set point exceeded
controller / LS 5
yes
no
no
E073
Temperature 1, table
value undershot
Check temperature value for plausibility;
if necessary, adjust or extend table.
yes
no
no
E075
Temperature 1, table
value exceeded
yes
no
no
Error
No.
N
A
M
U
R
Error message
Possible causes / measures
Alarm
contact
Error
current
Automatic
cleaning
start
Fact.
User
Fact.
User
Fact.
User