6 Commissioning
TopClean S CPC 30
34
EHauser
6.5
Description of functions
In this item, you change the settings for measured value acquisition, such as the
operating mode, the measuring principle, the electrode type etc.
Apart from the signal attenuation, you have already made all the settings in the menu at
the first commissioning in the Quick Setup (s. page 31). You can change the selected
values in this menu.
PARAM
⇒
Set up 1
⇒
Sensor input
CODE
CHOICE
(default = bold)
INFO
(E1, 2 = Editor types, s. page 28)
User
settings
A1
pH
Redox mV
Redox %
Select operating mode
If the operating mode changes, the user settings are
automatically reset.
E1
A4
Glass el.
Glass el.
IsFET
Antimony
7.0
4.6
4.6
Select electrode type
(only pH)
E1
A6
Solution ground
No solution ground
Select connection type
Solution ground
= symmetrical = with potential
matching (PML)
No solution ground
= unsymmetrical = without PML
!
Note!
For further information, see page 23
E1
A7
pH/Redox:
Temperature:
(01 ... 30s)
01 s
01 s
Set measured value attenuation
The mean value over the set time is displayed.
00s = no attenuation
E2
PARAM
⇒
Set up 1
⇒
Display
CODE
CHOICE
(default = bold)
INFO
(E1, 2 = Editor types, s. page 28)
User
settings
B1
E
D
F
I
ES
Select language
Depending on ordered language version:
Language version variant A:
E / D / F / I / ES
Language version variant B:
E / D / NL / J
E1
B2
Contrast setting as necessary
You can increase and reduce the contrast with the
+/- keys.
–
Hold
pH
7.00
Param
Edit
+-
Contrast
Next E