TopClean S CPC 30
4 Wiring
Hauser
19
4.3.2
Contact assignment Mycom S
In the transmitter Mycom S CPM 153, there are six relays available which you can
configure via user menu (see the "PARAM" menu
➞
"Set up 1"
➞
"Contacts" from
page 38).
!
Note!
•
IIf you use NAMUR contacts (acc. to recommendations of process control
engineering of the chemical and pharmaceutical industry), the contacts are set to the
relays as follows:
– "Failure" to ALARM
– "Maintenance required" to RELAY 1 and
– "Function check" to "RELAY 2".
•
You can assign up to three relays to the controllers.
Besides the automatic cleaning / calibration you can realize additional cleaning
functions via the
ChemoClean
®
relays. This increases the flexible use of TopClean S.
You can switch the contact type NC / NO via software.
Selection by software
NAMUR
on
NAMUR
off
ALARM
Failure
Alarm
RELAY 1
Warning when
maintenance required
freely selectable
RELAY 2
Function check
freely selectable