TopClean S CPC 30
6 Commissioning
Hauser
49
PARAM
⇒
Set up 2
⇒
Check systems
CODE
CHOICE
(default = bold)
INFO
(E1, 2 = Editor types, s. page 28)
User
settings
L1
SCS:
SCS Ref.:
off
light
Select SCS (= Sensor Check System) mode:
SCS
: Recognition of glass breakage
(off;
Ref
. = reference electr.; Glass = measuring
electrode;
G+R= measuring reference electr.)
SCS Ref
.: Blockage recognition
(light, medium, heavy, very heavy blockage)
!
Note!
In a unsymmetrical connection (without PML) only the
glass electrode can be monitored.
E2
L2
PCS:
off
PCS (= Process check system) time
If the measuring signal does not change during the
entered time for ±0.02 pH / ±5mV / ±0.25%, an alarm is
signalled with error message E152.
Settable times: off, 1h, 2h, 4h.
!
Note!
An active PCS alarm signal will be deleted
automatically as soon as the sensor signal changes.
E2