8 Trouble-shooting
TopClean S CPC 30
102
EHauser
8.1.2
Process-specific errors
Error
Possible cause
Remedial action
Equipment needed, spare
parts
Instrument
unconfigurable,
Display for code prompt
is 9999
Instrument hardware is locked via
keypad (Keys "CAL" + "DIAG"
simultaneously = security locking)
Press "MEAS" and "PARAM"
simultaneously to unlock.
No or slow
display change
Electrode soiled
Clean electrode.
Electrode aged
Replace electrode.
New electrode
Electrode defective (reference lead)
Replace electrode.
New electrode
Inner buffer missing
Check KCl supply (0.8 bar above
medium pressure).
KCl (CPY 4-x)
Problem with diaphragm or
missing electrolyte
Check KCl supply (0.8 bar above
medium pressure).
KCl (CPY 4-x)
No calibration possible
as sensor adjustment
time too long.
For IsFET sensor: Moisture film of
measuring surface interrupted due
to drying or blowing out with
compressed air.
Ensure moisture film is present or
ensure buffer dwell time > 6 min.
Do not clean IsFET sensor with
compressed air.
Permanent, incorrect
measured value
Electrode not immersing or
protection cap not removed
Check installation position, remove
protection cap.
Air cushion in assembly
Check assembly and installation
position.
Earth fault at or in the instrument
Test measurement in insulated vessel,
possibly with buffer solution.
Plastic vessel, buffer solutions.
Behaviour when instrument is
connected to process?
Hairline crack in glass membrane
Replace electrode.
pH electrode
Instrument in impermissible
operating status (no response on
pressing key)
Switch instrument off and on.
EMC problem: If repeated, check
grounding and wire routing
Incorrect temperature
reading
Incorrect sensor connection
Check connections using wiring
diagram.
Wiring diagram s. page 15
Measuring cable defective
Check cable.
Ohmmeter
Incorrect sensor type selected
Set sensor type on instrument
(Field 141).
Check temperature sensor with
Ohmmeter.
Sensor defecitve
Check sensor
pH value in process
incorrect
No / incorrect temperature
compensation
ATC: Activate function
MTC: Set process temperature
Conductivity of medium too low
Select pH electrode with salt supply
or liquid KCl.
e.g. Orbisint CPS 11-xASxx
or Ceraliquid CPS 41
Flow rate too high
Reduce flow rate or measure in a
bypass.
Potential in medium
Possibly earth with / at PA pin (connect
PA to PE).
Problem mainly occurs in plastic lines
Device unsymmetrical and PML
connected
Possibly earth with / at PA pin (connect
PA to PE).
Electrode soiled or with deposits
Clean electrode (s. Chap. 8.8.1).
Highly polluted media:
Use spray cleaning.