6 Commissioning
TopClean S CPC 30
60
EHauser
ME2
Set point
St. ntr. zone
End ntr. zone
K
R
1
K
R
2
pH 07.00
pH 06.50
pH 07.50
pH 01.00
pH 01.00
Characteristic values for constant characteristic
curve (constant control gain)
Set point
: The value which should be set.
St. ntr. zone:
Start neutral zone
End ntr. zone: End neutral zone
K
R
1
(only with alkali dosing): Modulation gain for alkali
dosing
K
R
2
(only with acid dosing): Modulation gain for acid
dosing
ME3
Set point
St. ntr. zone
End ntr. zone
O.pnt. X1
O. pnt.Y1
O.pnt. X2
O. pnt.Y2
Ctrl.pnt.1
Ctrl.pnt. 2
pH 07.00
pH 06.50
pH 07.50
pH 05.00
0.20
pH 09.00
-0.20
pH 02.00
pH 12.00
Characteristic values for segmented characteristic
curve
(range-dependent control gain)
Set point
: The value which should be set.
St. ntr. zone:
Start neutral zone
End ntr. zone:
End neutral zone
Optimisation point 1 and 2:
Entry with x and y
coordinates
Control point 1
: The dosing is 100 % alkali for measured
values < control point 1.
Control point 2
: The dosing is 100 % acid for measured
values > control point 2.
ME4
Rapid process
Standard process
Slow process
User settings
Select process character
If you have no experience in setting control
parameters, these defaults
rapid
/
standard
/
slow
process
are intended as an aid to adapting the
controller behaviour to the process. Select a default
and use the "controller simulation" (see below) to check
if these settings are relevant for your process.
Enter all the characteristic values yourself with the
user
settings
.
E1
ME5
K
R
1 =
K
R
2 =
Tn 1 =
Tn 2 =
Tv 1 =
Tv 2 =
Characteristic values for user settings:
(
K
R
1 and
K
R
2 only with linear characteristic curve;
Index 1 only for alkali dosing, Index 2 only for acid
dosing)
K
R
1
: modulation gain for alkali dosing
K
R
2
: modulation gain for acid dosing
Tn
:
integral action time
Tv
: derivative action time
E2
ME6
Simulation
Off
On
Select controller simulation
Here, you can switch a configuration loop on or off. The
hold is removed with an active controller simulation.
Simulation on
: The characteristic values entered in the
previous field are used in the next field to simulate the
controller behaviour.
Off
: Press "E" to leave the controller simulation.
E1
ME7
Function
Set:
Act.:
Y:
auto
pH 07.00
pH 07.00
000
Controller simulation
Function
: Here, you set whether an actuating variable
calculated by the controller ("auto"), or an actuating
variable y entered by the user ("manual") is to be
output.
Set
: Displays the current set point. If necessary, you
can change the set point. The other points (start/end of
neutral zone, optimization points, control points)
change accordingly.
Actual
: Displays the current actual/measured value.
Y
: With the "auto" function: displays the actuating
variable determined by the controller. With the "manual"
function, you can enter an actuating variable here.
Values < 0 % mean a dosing of acid, values > 0 %
mean a dosing of alkali.
E2
CODE
CHOICE
(default = bold)
INFO
(E1, 2 = Editor types, s. page 28)
User
settings