TopClean S CPC 30
6 Commissioning
Hauser
41
Medium compensation (only for pH):
GB1
Select table
Create tables
Reference temperature
Selection
Enter / activate customer-specific temperature
compensation tables.
Select table
= Select for activation
E1
Select table:
GBA1
Medium 1
Medium 2
Medium 3
Off
Select medium
Select a medium.
Off
= no medium compensation
E1
Create tables:
GBB1
Medium 1
Medium 2
Medium 3
Select medium
Medium compensation curves can be entered
as a table for three different media.
E1
GBB2
02
(2 ... 10)
Entry of the number of support points (value pairs)
Value pair: pH/redox and temperature
E2
GBB3
°C
020.0 °C
025.0°C
pH
02.00
04.00
Entry of value pair
Enter pH/redox and temperature (number of required
value pairs = number of support points desired in Field
GBB2).
E2
GBB4
OK
Delete element(s)
Selection:
Are the value pairs OK or do you want to delete
elements?
E1
GBB5
°C
020.0 °C
025.0°C
pH
02.00
04.00
Delete:
Select the rows to be deleted, delete them with
T
and
confirm this with "E".
E2
GBB6
Valid table
Note in display:
The table is active after confirmation by pressing "E".
Cancel by pressing "PARAM".
–
Reference temperature:
GBC1
For laboratory
measurement:
25.0 °C
(0 ... 100 °C)
Entry of reference temperature
to which the medium temperature shall be
compensated. Here, enter the temperature at which the
pH set point of the process is defined, (e.g. the
ambient temperature in the laboratory).
E2
CODE
CHOICE
(default = bold)
INFO
(E1, 2 = Editor types, s. page 28)