6 Commissioning
TopClean S CPC 30
32
EHauser
T10
°C
°F
Select temperature display
T11
ATC
MTC
MTC+Temp
Select temperature compensation
ATC
= automatic temperature
compensation
MTC
= manual temp. comp. (with fixed
temperature, entered in Field GAA2
(page 40).
MTC+Temp
. = as MTC. On the display
however, the value which appears is
what the temperature sensor measures in
the medium.
T12
025.0 °C
Temperature value
(only with pH and selection of MTC or
MTC+Temp in Field T11)
T13
Off
On
Temperature measurement
(only for redox)
T17
Pt 100
Pt 1000
NTC 30k
Select temperature sensor
T18
NAMUR
Relay 1
Relay 2
off
free
free
Contact functions
Depending on the equipment available,
you can assign the function of up to 5
relays here. The relays 1 and 2 will be
assigned to an activated NAMUR
function and will not be available for
other functions (compare p. 34).
Selection:
Free / Controller / LC / CCW / CCC
Controller:
Controller control using relay
LC
: Limit contactor function
CCW
: ChemoClean water. Water supply
for the ChemoClean function.
CCC
: ChemoClean Cleaner. Cleaner
supply for the ChemoClean function.
(Together, CCC and CCW form the
"ChemoClean" function. You can find
information on ChemoClean on page 71).
CODE
Display
CHOICE
(default = bold)
INFO
User settings
Hold
pH
7.00
Param
Edit
E
°C
°F
Temp. unit
pH
7.00
Param
Edit
ATC K1
MTC
MTC+Temp
E
Temp.conf. K1
Next
Hold
pH
7.00
Param
Edit
-20.0...150.0°C
E
0
25.0
MTC-Temp. K1
Next
Hold
mV -
114
Param
Edit
E
Temp.meas1
Next
on
off
pH
7.00
Param
Pt100
Pt1000
NTC30k
Edit
Hold
E
Temp.sensor
Next
Hold
pH
7.00
Param
E
Relay funct.
Next
Acc.Namur
off
Select
Relais 1
free
Relais 2
free