6 Commissioning
TopClean S CPC 30
72
EHauser
Operation:
1.
In the menu "Set up 1"
➞
"Contacts" (Field F1, s. page 38), the function
ChemoClean
®
must be switched on and the appropriate contacts connected to the
injector, (see connection examples on page 136 and page 137).
2.
The cleaning processes are configured in this menu. Here, the automatic or event-
controlled cleaning can be adapted to the process conditions.
One or more of the following controls are possible:
– Weekly programme (see below, Fields OA1 to OAA5): Any number of cleanings
can be started on each weekday
– External control: a start can be triggered via the digital input. For this, external
control must be activated in Field O1, "Select control levels": Ext. control "on")
– Clean trigger: Cleaning is carried out when an SCS alarm occurs (see also Field
LM1 under "Set up 2"
➞
"Check systems")
– Power failure: cleaning is started after a power failure.
3.
Rapid on-site cleaning can be carried out with the menu:
Manual operation:
Quick on-site cleaning can be carried out using the menu:
"PARAM"
➞
"Manual operation"
➞
"ChemoClean"
➞
press "E" twice ("start cleaning")
Weekly programme:
"PARAM"
➞
"Set up 2"
➞
"ChemoClean":
Each day can be programmed individually. The following programmes are available
•
"Clean": Cleaning triggered by entering the start time (s. fig. 30).
•
"Clean Int": Cleaning is carried out at intervals with a defined spacing (s. fig. 30). This
programme cannot be started directly via the binary inputs.
•
"User": User-defined cleaning programmes (create in Programme Editor; from Field
NAD1).
Programme sequences (cleaning example)
Monday:
2 x cleaning (at 11:00 and at 18:00) with 120 s. water, of which 60 s. additionally with
cleaner.
Clean every 30 mins. between 18:20 and 24:00 (= 1800 s.) with 120 s. water, of which
60 s. additionally with cleaner.
Fig. 30: Graphic representation of the above example
C07-CPM153xx-05-06-00-en-001.EPS
Clean
Hold
Hold and Hold delay time
water
60 sec
60 sec
+cleaner
+cleaner
Clean Int
Hold
Hold and Hold delay time
Hold and Hold delay time
water
60 sec
60 sec
60 sec
60 sec
Meas. time1800 sec
+cleaner
+cleaner
+cleaner
Measuring
time