TopClean S CPC 30
3 Installation
Hauser
11
3.3
Installation
3.3.1
Wall mounting
"
Caution!
•
Check that the temperature does not exceed the maximum permitted ambient
temperature (–20° ... +60°C). Install the devices in a shady location. Avoid direct
sunlight.
•
Mount the devices so that the cable entries always point downwards.
Control unit
CPG 30
Fig. 3:
Mounting the control unit CPG 30 using wall mounting kit (contained in scope of delivery)
For the wall mounting of the control unit CPG 30 proceed as follows:
1.
Prepare drill holes acc. to fig. 3.
2.
Screw the elements of the supplied wall fixing kit to the back panel of the housing
(screws contained in the scope of delivery).
3.
Fix the housing to the wall.
Transmitter
Mycom S CPM 153
Fig. 4:
Dimensions for the wall mounting:
Fixing screw:
dia.
6 mm
Wall plug: dia. 8 mm
1: Fixing drillholes
2: Plastic cover caps
C07-CPC300xx-06-12-00-xx-001.EPS
320
250
227
327
350
420
154
100
247
142
167
∅
8
2
C07-CPM153xx-11-00-08-xx-001.EPS
1