6 Commissioning
TopClean S CPC 30
46
EHauser
Special buffer:
JB1
2
(2 ... 3)
Entry of the number of buffers
In a table, you can save up to 3 buffers not saved in the
instrument.
!
Note!
Fields JB2 to JB6 must be run through individually for
each buffer.
E2
JB2
1
(1 ... 2)
Edit table
Select one of the tables for editing
E2
JB3
10
(2 ... 10)
Entry of the number of support points (value pairs)
Value pair: pH and temperature
E2
JB4
°C:
000.0
005.0
...
pH:
04.00
04.05
...
Entry of value pair
Enter temperature and pH/Redox (number of required
value pairs = number of support points desired in
Field JB3).
E2
JB5
OK
Delete element(s)
Selection:
Are the value pairs OK or do you want
to delete any of them?
E1
JB6
°C:
000.0
005.0
...
pH:
04.00
04.05
...
Delete:
Select the rows to be deleted, delete them with
T
and
confirm this with "E".
E2
JB7
Valid table
Note in display:
The table is active after confirmation by pressing "E".
Cancel by pressing "PARAM".
–
General settings:
JC1
ATC 1
MTC
Select the temperature compensation for the
calibration
ATC = Automatic temp. comp.
MTC = Manual temp. comp.
!
Note!
The setting is only active during calibration. In
operation, the setting selected in GAA1 is valid.
E1
JC2
45.00 mV/pH
(5.00 ... 99.00 mV/pH)
Entry of slope for alarm function
If the entered slope is not reached, an alarm (error no.
032 / E035) can be triggered (error activation in
Field H5).
E2
JC3
pH 2.00
(0.05 ... 2.00 pH)
Entry of pH value zero point deviation
for the alarm function
If the zero point deviates from the reference zero point
by the value entered here, an alarm (Error no. 033) can
be triggered (error activation in Field H5).
E2
CODE
CHOICE
(default = bold)
INFO
(E1, 2 = Editor types, s. page 28)
User
settings