Commissioning
94
Optional program cycles
* For redox measurement, the "RedoxCal" program is available instead of the "User 1" program (see
the following page).
User 1 - User2*
Val. P1
Val. P2
Val. P1/2
Val+Cal.
01
01
Assembly
service
01
Assembly
service
01
Assembly
service
01
Assembly
service
02
02
Water
60 s
02
Water
60 s
02
Water
60 s
02
Water
60 s
03
03
Cleaner
3s
03
Cleaner
3s
03
Cleaner
3s
03
Cleaner
3s
04
04
Wait
120s
04
Wait
120s
04
Wait
120s
04
Wait
120s
05
05
Water
60 s
05
Water
60 s
05
Water
60 s
05
Water
60 s
06
06
Compressed
air
20s
06
Compressed
air
20s
06
Compressed
air
20s
06
Compressed
air
20s
07
07
Back to 2
1x
07
Back to 2
1x
07
Back to 2
1x
07
Back to 2
1x
08
08
Pump buffer 1 3s
08
Pump buffer 2
3s
08
Pump buffer 1 3s
08
Pump buffer 1 3s
09
09
Wait
60 s
09
Wait
60 s
09
Wait
60 s
09
Wait
60 s
10
10
Val. buffer 1
10
Val. buffer 2
10
Val. buffer 1
10
Val. buffer 1
11
11
Water
60 s
11
Water
60 s
11
Water
60 s
11
Cal. Buffer 1
12
12
Compressed
air
20s
12
Compressed
air
20s
12
Compressed
air
20s
12
Water
60 s
13
13
Assembly
measuring
13
Assembly
measuring
13
Pump buffer 2 3s
13
Compressed
air
20s
14
14
14
14
Wait
60 s
14
Pump buffer 2 3s
15
15
15
15
Val. buffer 2
15
Wait
60 s
16
16
16
16
Water
60 s
16
Val. buffer 2
17
17
17
17
Compressed
air
20s
17
Cal. Buffer 2
18
18
18
18
Assembly
measuring
18
Water
60 s
19
19
19
19
19
Compressed
air
20s
20
20
20
20
20
Assembly
measuring
21
(up to 28 program
steps possible)
21
21
21
21
22
22
22
22
22
Содержание Topcal S CPC310