Medium connection
46
Connecting pneumatic assembly control system
Connect the connections for moving the assembly and for position confirmation as follows:
With inductive limit position switches
a0005752
Fig. 49:
Connecting CPA471, CPA472, CPA475 compressed air control with inductive limit position switches
Connecting pneumatic assembly control system
Connect the connections for moving the assembly as follows:
Hose
number
Function
Assembly connection
5
"Measure" position confirmation
Lower limit position switch - output (=2)
2
Move to "Measuring" position
Lower limit position switch - input (=1) and upper G¼ coupling (via
T-piece)
6
"Service" position confirmation
Upper limit position switch - output (=2)
3
Move to "Service" position
Upper limit position switch - input (=1) and lower G¼ coupling (via
T-piece)
3
2
2
3
M1
Hose
number
Function
Assembly connection
2
Move to "Measuring" position
Upper G¼ coupling
3
Move to "Service" position
Lower G¼ coupling