Commissioning
58
7.4
Switching on
Before first startup, make sure you understand how to operate the device. Please refer to the "Safety
instructions" and "Operation" sections in particular.
We recommend the following procedure for commissioning:
1.
Connect Mycom S CPM153 to the power supply.
2.
Switch the service switch to the Service position
.
3.
Wait for CPM153 and the control unit to initialize.
Function of the green "Alive LED":
– Frequency of approx. 2 pulses per second: Communication is active.
– Frequency of approx. 1 pulse per second: Establishing communication.
– LED lit constantly: No communication.
If the LED is not lit, check the power supply for terminal L+/L– (12 ... 15 V DC).
4.
Only initial commissioning:
Run the "Quick Setup" (see "Quick Setup" section).
5.
Switch the service switch to the Measure position
.
6.
Configure parameters: Select a function for the additional valves (optional).
7.
Start the quick-test program "User 3" and check the entire system for leaks.
Program start:
"PARAM > Specialist > Manual operation > Topcal > Status messages (Enter) > Start program
> User 3".
8.
Then configure the system completely via the software.
First commissioning
On first switch-on, the instrument starts automatically with the Quick Setup menu. This asks you
about the most important instrument settings. After you close the menu, the instrument is ready for
use and measurement in its standard configuration.
!
Note!
• You must completely run through the Quick Setup menu. If you do not, the instrument will not
be operational. If you interrupt Quick Setup, it will start again the next time you start it up until
all
the items have been processed and completed.
• You have to enter the specialist access code (factory setting 0000) to configure.
7.5
Quick Setup
The Quick Setup menu allows you to configure the most important functions of the transmitter. It
is started automatically during initial commissioning and can be called any time via the menu
structure.
To call the menu, select:
Ñ
> Specialist > Spec. access code: 0000 > First start up
Function
Options
Info
Language
Options
• English
• Language ordered
Factory setting
English
!
Note!
""Language ordered" refers to the language that
you chose for your device by means of the order
code ("Language version").
All the other settings remain if the other
language is selected.
pH 7. 00
Hold
Param
Language
English
GB
Deutsch
D
Edit [¯]
Next [E]