Wiring
37
External outputs
1.
If you want to report the position of the assembly to an external PLC, connect the outputs of
the control unit as follows:
4.1.11
External inputs (PLC to Mycom)
a0003970
Fig. 41:
Connecting external inputs
1.
If you want to activate the hold function for Mycom by means of an external PLC, connect the
input to terminals 81 and 82 of the device (power supply necessary).
2.
If you want to control the Chemoclean programs by means of an external PLC, connect the
inputs as follows:
Feedback
Control unit connection
"Assembly in Measuring position"
feedback signal
Terminals 61 and 62
"Assembly in Service position"
feedback signal
Terminals 65 and 66
85 81 93 89
86 82 94 90
- +
- +
- +
Chemoclean program
Mycom connection
"Clean" program
Terminals 93 and 94
"User" program
Terminals 89 and 90