Commissioning
95
Programs for redox operating mode
You cannot calibrate with the Clean C and Clean CS calibration programs in the redox operating
mode. Instead, you can use the "Redox Cal." program in User program 1.
Configuration menu
To call the menu, select:
"
Ñ
> Specialist > Set up 2 > Topcal"
RedoxCal
01
Assembly service
02
Water
60 s
03
Cleaner
3s
04
Wait
120s
05
Water
60 s
06
Compressed air
20s
07
Back to 2
1x
08
Pump buffer 1
3s
09
Wait
60 s
10
Cal. Buffer 1
15s
11
Water
60 s
12
Compressed air
20s
13
Assembly measuring
14
15
(up to 28 program steps
possible)
Function
Options
Info
Select function
Options
• Setup Topcal
• Activate Topcal
Factory setting
Setup Topcal
Setup:
Create or edit a Topcal program.
Activate:
Switch Topcal functions on or off.
Configuration
Note
Automatic: Off
Clean trigger: Off
Ext. control: Off
Current system status
Valve name V1 (or V2)
0 ... 9; A ... Z
Factory setting
Valve 1 (or 2)
You can enter names up to eight characters in
length for the additional valves.
These names are automatically accepted with
the program steps.
pH 7.00
Hold
Param
Topcal
Set up Topcal
Activate Topcal
Edit [¯]
Next [E]