Installation
14
3.4
Installation instructions
3.4.1
Securing rinsing block to assembly
a0006202
Fig. 8:
Mounting the rinsing block on the assembly (example CPA473)
Proceed as follows to mount the rinsing block:
1.
Fit the securing clamps with the rinsing block (3 and 4) on the assembly cylinder.
2.
Fit the counterclamps (2) on the assembly cylinder from the other side.
3.
Connect the clamps with the screws (1) supplied.
3.4.2
Installation instructions
• The Mycom S transmitter is used as a field device as standard. It can also be installed as a
panel-mounted instrument.
• Mycom S is suitable for wall mounting with securing screws and for post mounting to cylindrical
pipes.
• Always install the transmitter horizontally in such a way that the cable entries are always pointing
downwards.
1
3
4
2