Troubleshooting
136
10.2
System error messages
You can read out the extended device status or current error messages via command 48. The
command delivers bit-encoded information.
Byte
NAMUR
Bit
Error no.
Short error description
Failure
0
E000
Not used
Failure
1
E001
Memory error
Failure
2
E002
Data error in EEPROM
Failure
3
E003
Invalid configuration
0
Failure
4
E004
Invalid hardware ID
Failure
5
E005
Unknown CPG code
Failure
7
E007
Malfunction of transmitter 1
Failure
0
E008
SCS message sensor 1 / ISFET 1 (ISFET leak current > 400 nA)
Failure
1
E009
Not used
Failure
2
E010
Temperature sensor 1 defective
1
Failure
3
E011
Not used
Failure
4
E012
CPC communication failure
Failure
5
E013
Assembly has not reached service position
Failure
6
E014
Assembly has not reached measuring position
Failure
7
E015
Not used
Failure
0
E016
Not used
Failure
1
E017
Data error in CPC EEPROM
Failure
2
E018
Not used
2
Failure
3
E019
Not used
Failure
4
E020
Not used
Failure
5
E021
Not used
Failure
6
E022
Not used
Failure
7
E023
Not used
Failure
0
E024
CPC program interrupted
Failure
1
E025
Not used
Failure
2
E026
Not used
Failure
3
E027
Compressed air failure
3
Failure
4
E028
Not used
Failure
5
E029
Sensor 1 self-test error (digital sensor)
Maintenance
6
E030
SCS message reference electrode 1
Maintenance
7
E031
Not used
Maintenance
0
E032
Outside set slope range for sensor 1
Maintenance
1
E033
Outside set zero point for sensor 1
Maintenance
2
E034
Outside set offset range for sensor 1
Maintenance
3
E035
Outside set slope range for sensor 2
4
Maintenance
4
E036
Not used
Maintenance
5
E037
Not used
Maintenance
6
E038
Not used
Maintenance
7
E039
Not used
Содержание Topcal S CPC310