Commissioning
64
7.6.2
Set up 1 - Display
To call the menu, select:
"
Ñ
> Specialist > Set up 1 > Display"
7.6.3
Set up 1 - Access codes
To call the menu, select:
"
Ñ
> Specialist > Set up 1 > Access codes"
Function
Options
Info
Language
Options
• English
• Language ordered
Factory setting
English
!
Note!
""Language ordered" refers to the language that
you chose for your device by means of the order
code ("Language version").
All the other settings remain if the other
language is selected.
Contrast
Contrast setting where applicable
You can increase or reduce the contrast of the
display with the arrow keys
V
and
W
.
Date + time
Weekday
Day
Month
Year
Time
Enter the complete date and time here.
These data are used for the logbooks and the
automatic cleaning system.
Number of decimal places
Options
• pH 00.00
• pH 00.0
Factory setting
pH 00.00
Only available for pH operating mode:
Specify whether the measured values are to be
displayed to one or two decimal places.
Temperature unit
Options
•
°
C
•
°
F
Factory setting
°
C
Tag number
0 ... 9; A ... Z
Enter your client-specific device number here
(32-digit tag number): This is also saved on the
DAT module which is available as an option.
Function
Options
Info
Service code
0000 ... 9997
Factory setting
0000
Enter the operator access code. This code
allows access to the calibration menu and the
temperature compensation menu item.
0000 = No locking
Specialist code
0000 ... 9997
Factory setting
0000
Enter the specialist code. This code allows
access to all the menu items.
0000 = No locking
pH 7. 00
Hold
Param
Language
English
GB
Deutsch
D
Edit [¯]
Next [E]
Hold
pH
7. 00
Param
Edit
[+-]
[E]
Contrast
Next
pH 7.00
Hold
Param
Service Code
0...9997
Edit [ ¯®]
Next [E]
0
000