Wiring
25
1.
Guide the cables of the level probes for the buffer and cleaner through the threefold Pg gland
behind the multihose connection (see Fig. 22).
2.
Connect the cable cores as follows. Polarity does not matter here:
4.1.6
Analog sensors
Measuring cables
You require shielded special measuring cables to connect pH and redox sensors to the transmitter.
You can use the following multicore and preterminated cable types:
!
Note!
Further information on the cables and junction boxes can be found in the "Accessories" section.
a0006021
Fig. 22:
Guiding the level probe cable
1
Threefold Pg gland
Cable core
Control unit connection
Level probe, buffer 1
Terminal P1 and P2
Level probe, buffer 2
Terminal P3 and P4
Level probe, cleaner
Terminal P5 and P6
Sensor type
Cable
Extension
Electrode without temperature sensor
CPK1
VBA / VBM box + CYK71 cable
Electrode with Pt 100 temperature sensor and
TOP68 plug-in head
CPK9
VBA / VBM box + CYK71 cable
ISFET sensor with Pt 100 / Pt 1000 temperature
sensor and TOP68 plug-in head
CPK12
VBA / VBM box + CYK12 cable
pH individual electrode with separate reference
electrode and separate temperature sensor
CPK2
VBA / VBM box + PMK cable