Commissioning
79
7.6.11
Set up 2 - Data log
The data logger records two freely selectable parameters with their date and time. You can call it up
by means of the measured value displays. Use the arrow keys to scroll through the measured value
displays until you get to the record mode of the data logger. Press the key to enter the scroll mode
of the data logger. Here you can open the saved measured values with their date and time.
To call the menu, select:
"
Ñ
> Specialist > Set up 2 > Data log"
Function
Options
Info
Data logger
Options
• Sample time
• Data log 1
• Data log 2
• DataLog display 1
• DataLog display 2
Factory setting
Sample time
With the data loggers you can
• record a parameter with 500 consecutive
measuring points.
• two parameters each with 500 sequential
measuring points.
Sample time
Enter sample time
2 ... 36000 s
Factory setting
00005 s
Enter the time interval after which the next
measured value is recorded in the data logger.
Data logger
Data log 1 (or 2)
Activation and subsequent entry
• Measured value
– pH/mV
– Temp
• Function
– On
– Off
Factory setting
Input: pH/mV
Function: Off
Select the measured value to be recorded and
activate the function with "On".
Recording range
Activation and subsequent entry
• Min
-2.00 ... 16.00 pH
-1500 ... 1500 mV
-50 ... 150
°
C
• Max
-2.00 ... 16 pH
-1500 ... 1500 mV
-50 ... 150
°
C
Factory setting
Min: 2.00 pH
Max. 12.00 pH
Specify the recording range. Values outside the
range defined here are not recorded.
DataLog display
View of the recorded data
You can call up the data recorded in the past
with date and time.
Hold
pH 7. 00
Param
[¯]
[E]
Data log
Sample time
Data log 1
DataLog display 2
Data log 2
Next
DataLog display 1
Edit
7.54 pH
12:15:35
pH 7.54
09.04.04
Para
DataLog View 1