Commissioning
60
MTC temperature, temp. sensor
input 1
-20.0 ... 150.0
°
C
Factory setting
25.0
°
C
Only available with pH and MTC or
MTC+Temp selected in the previous field.
Temperature measurement
Options
• Off
• On
Factory setting
On
Only available with redox measurement.
Contact functions
Activation and subsequent entry
• NAMUR
– On
– Off
• Relay 1-5
– N/C
– Controller
– Limit
– CCW
– CCC
Factory setting
NAMUR: Off
Relay 1-5: N/C
You can specify the function of the five relays
here.
When you switch on NAMUR, relays 1 and 2
are assigned and are not available for another
function (see "Namur assignment" section).
Controller:
Relay contact for controller output
Limit:
Limit switch function
CCW:
Chemoclean water. Supply of water for
the Chemoclean function.
CCC:
Chemoclean cleaner. Supply of cleaner
for the Chemoclean function.
(Together, CCC and CCW form the
"Chemoclean" function; information on
Chemoclean is provided in the "Set up 2 -
Chemoclean" section.)
Current output 1, output variable
Options
• pH/mV Input 1
• Temperature Input 1
Factory setting
pH/mV Input 1
Select the measured value to be output at
current output 1.
Current output 2, output variable
Options
• pH/mV Input 1
• Temperature Input 1
• Continuous controller
Factory setting
Temperature Input 1
Select the measured value to be output at
current output 2.
Continuous controller:
The controller
manipulated variable is output via the current
output (see also Controller menu).
!
Note!
Danger of data loss. If you change the
assignment for the current output from
"continuous controller" to another function
once you have configured the controller, the
entire controller configuration is reset to the
default values.
Tag number
0 ... 9; A ... Z
Enter your client-specific device number here
(32-digit tag number): This is also saved on the
DAT module which is available as an option.
Function
Options
Info
Hold
pH
7. 00
Param
Edit
-20. 0... 150. 0°C
0
25. 0°C
[¯ ®]
[E]
MTC-Temp. C1
Next
Hold
mV -
114
Param
Edit
[¯]
[E]
Temp.meas1
Next
on
off
Hold
pH
7. 00
Param
[¯ ®]
[E]
Relay funct.
Next
Acc.Namur
off
Select
Relay 1
N/C
Relay 2
N/C
Relay 3
N/C
¯Relay 4
N/C
Hold
pH 7.00
Param
Output 1
pH/mV Input 1
Temperature Input 1
Edit [¯]
Next [E]
Hold
pH
7. 00
Param
Edit
[¯]
[E]
Output 2
pH/mV Input 1
Temperature Input1
Controller
Next
pH
7. 00
Param
Edit
Hold
0...9, A...z
[¯ ®]
[E]
Tag number
Next