Troubleshooting
145
Behavior with NAMUR setting (contacts configured as normally closed contacts)
10.5.3
Response of contacts to power failure
The contacts can be defined as normally open or normally closed contacts in the "Set up 1 > Relays"
menu (see "Set up 1 - Relays" section). In the case of a power failure, the contacts will act according
to the setting you make.
10.5.4
Assembly behavior
10.6
Spare parts
Please order spare parts from your local sales center. For this purpose, use the order numbers listed
in the "Spare part kits" section.
You should
always
quote the following information on the spare parts order:
• Device order code
• Serial no.
• Software version if possible
You can refer to the nameplate for the order code and serial number.
The software version is provided in the device software (see "Operation" section) provided the
processor system of the device is still working.
Device status
Alarm relay
Maintenance
relay
Function check
Limit value / Controller
Normal operation
Picked up (failsafe
behavior)
Picked up
Picked up
Appropriate configuration
and operating status
Failure
Dropped out
Picked up
Picked up
Appropriate configuration
and operating status
Maintenance required
Picked up
Dropped out
Picked up
Appropriate configuration
and operating status
Function check
Picked up
Picked up
Dropped out
Appropriate configuration
and operating status
Voltage-free
Dropped out
Dropped out
Dropped out
Dropped out
Problem
CPM153 behavior
CPG behavior
Assembly behavior
CPM153 voltage failure during
measurement
No function
CPM153 voltage failure during
maintenance
No function
CPG voltage failure during
measurement
Error message E012
All valves closed
Can be pressed from the process
CPG voltage failure during
maintenance
Error message E012
All valves closed
Stays in service position
Air failure during measurement
Error message E027
Can be pressed from the process
Air failure during maintenance
Error message E027
Stays in service position
Air failure during automatic
cleaning
Error message E027
Stays in service position
Buffer 1 / 2 empty
Error message E051/E052
Cleaner empty
Error message E050