Commissioning
66
Linear
Measured value limits
Activation and subsequent entry
• 0/4 mA
-02.00 ... 16.00 pH
-50 ... 150
°
C
-0500 ... 0500 mV
• 20 mA
-02.00 ... 16.00 pH
-50 ... 150
°
C
-0500 mV ... 0500 mV
Factory setting
0/4 mA: 02.00 pH / 000.0
°
C /
-0500 mV
20 mA: 12.00 pH / 100.0
°
C /
0500 mV
Enter the upper and lower measured value
limit.
The minimum distance between the upper and
lower measured value limit is 2 pH units (e.g.
0/4 mA: pH 7 and 20 mA: pH 9)
Note
Linear active.
The linear curve is active after confirming with
F
. Cancel with
Ñ
.
Table
Total pairs
01 ... 10
Specify the number of value pairs (table points)
for your table.
Table
Activation and subsequent entry
• pH/redox/temperature
-02.00 ... 16.00 pH
-0500 ... 0500 mV
-50 ... 150
°
C
• mA
0.00 ... 20.00 mA
Enter the necessary value pairs. The number of
value pairs corresponds to the value entered in
the previous field.
Example of value pairs with four table points:
a0004223
Confirm
Options
• OK
• Delete element
Factory setting
OK
Are the value pairs OK or do you want to delete
elements?
Delete:
In the screen that follows, select the
row to be deleted, delete with
T
and confirm
with
F
.
Table status notice
Valid table
Status of table. If invalid, then back to the
previous field.
Activate
Table active
The table is active after confirming with
F
.
Cancel with
Ñ
.
Function
Options
Info
8
4
2
6 8
12
20
mA
16
pH