Operation
53
6.1.3
Service switch
The service switch is located on the housing front of the control unit. Two switching positions are
available:
!
Note!
The service switch always has priority (Emergency Stop function). This means that any program
running is aborted as soon as the service switch is switched.
Ï
"CAL" takes you to the calibration menu to calibrate sensors.
F
With the "Enter" key, you proceed one step further in the menu or you
confirm the option selected.
LED lit
Green: Everything OK
Red: An error has occurred
VT
W
• You can use the arrow keys to scroll through the menu items and select
the option you require (if a choice is possible).
• Increase/decrease numbers by one level with "+" / "-". Go to the next
number with the "right arrow" (editor type 1) or
• ""Activate" with the "right arrow" and scroll through the option with "+"
/ "-" (editor type 2) (see the "Menu editor types" section on the editor
types)
Service/Off:
(horizontal switch setting)
• The sensor moves into the rinse chamber.
• "Hold" is active for the outputs.
Measure/On:
(vertical switch setting)
After moving out of the Service position, you are asked whether a program
should be started or whether the sensor should be moved into the process
without cleaning.
Only those programs are offered which have already been edited.