Appendix
170
12.2
Wiring example for external program start
a0005030
Fig. 80:
Power diagram for external control of the cleaning and calibration programs
1-8
Buttons to start the cleaning programs
81-86 Terminals for program start
0/1/2 Binary inputs of the CPG310 control unit
10-40V e.g. via power supply input of Mycom S CPM153, terminals 85/86 (15 V)
Diodes 1N4007
3 mA Per optoelectronic coupler input
1
2
3
4
6
7
8
1
2
0
82
bin 0
bin 1
bin 2
10 … 40 V
+
–
84
83
86
85
CPG310
81
User 3
User 2
User 1
Clean CS
Clean S
Clean C
Clean