Commissioning
89
Limit switch menu
To call the menu, select:
"
Ñ
> Specialist > Set up 2 > Limit switch"
Function
Options
Info
Limit switch
Options
• Limit switch 1
• Limit switch 2
• Limit switch 3
• Limit switch 4
• Limit switch 5
Select the limit switch you want to configure.
There are five limit contactors available.
Limit switch 1 ... 5
Activation and subsequent entry
• Function
– On
– Off
• Assignment
– pH/mV Input 1
– Temperature Input 1
• On value:
-2.00 ... 16.00 pH
-1500 ... 1500 mV
-3000 ... 3000 %
-50 ... 150
°
C
• Off value
-2.00 ... 16.00 pH
-1500 ... 1500 mV
-3000 ... 3000 %
-50 ... 150
°
C
Factory setting
Function: Off
Assignment: pH/mV
On-value: 16.00 pH
Off-value: 16.00 pH
Configure the limit switch.
Function:
Activate the function as a limit
switch
Assignment:
Select the measured value to
which the limit value should apply.
On-value:
Enter the value at which the limit
function is activated.
Off-value:
Enter the value at which the limit
function is deactivated.
Limit switch configuration
Activation and subsequent entry
• On delay
0 ... 2000 s
• Off delay
0 ... 2000 s
• Alarm lim.
-2.00 ... 16.00 pH
-1500 ... 1500 mV
-3000 ... 3000 %
-50 ... 150
°
C
Factory setting
On delay: 0s
Off delay: 0s
Alarm lim.: 16.00 pH
Configure the delays and the alarm threshold
for the limit switch.
On delay:
Enter the on delay
Off delay:
Enter the off delay
Alarm limit:
Enter the value at which the
fault-signaling contact switches.
Hold
pH 7. 00
Param
[E]
Next
Edit[¯]
Selection
Limit switch 1
Limit switch 2
Limit switch 3
Limit switch 4
Limit switch 5