Commissioning
109
7.8
Diagnosis
To call the menu, press
Ð
.
Function
Options
Info
Diagnosis
Options
• Error list
• Error log
• Operation log
• Calibration log
• Validate log
• Ext. sensor data (only for
digital sensors with
Memosens technology)
• Service
Error list:
Displays the error currently active.
(Complete error list with description, see the
"Troubleshooting" section).
Error log:
Lists the 30 errors last reported with
date and time.
Operation log:
Lists the 30 operating steps last
registered with date and time.
Calibration log:
Lists the last 30 calibrations
performed with date and time.
Validate log:
Lists the last 30 Topcal validati-
ons.
Ext. sensor data:
Lists the data saved in the
sensor, e.g. sensor identification, calibration
data, hours of operation etc.
!
Note!
• Use the arrow keys to scroll through the lists.
• Exit the lists with
F
.
Calibration log
Calibration data
• 1 data input
• Zero point
• Slope
• El. condition
• <Date> <time>
1 data input
Displays the calibration method
used.
Zeropoint:
Displays the zero point calculated
during calibration.
Slope:
Displays the slope calculated during
calibration.
Electr. condition:
Displays the electrode con-
dition.
<Date> <time>:
Displays the date and time of
the calibration.
If you are using a digital sensor with Memosens functionality, you get the following information after pressing
T
:
• SNR
• Sensor change date
• <Date> <time>
Displays the serial number of the calibrated
sensor and the date and time of the sensor
change.
Ext. sensor data (only for sensors with Memosens technology)
The transmitter indicates that the sensor data are being read out. The display automatically switches on completion of the
read-out process.
If the system does not automatically continue, you can call up the data last read out by pressing
F
or return to
measuring operation by pressing
Ò
.
Data of the digital sensor
Options
• Identification
• Calibration data
• Comp. temperature
• Sensor status
• Sensor info
Select the data saved in the digital sensor which
are to be displayed.
Identification
Manufacturer data
• ID
• SW ID
• HW version
• SW version
ID:
Displays the module ID of the sensor.
SW ID:
Displays the software ID of the sensor.
HW version:
Displays the hardware version of
the digital sensor.
SW version:
Displays the software version of
the digital sensor.
• Check date
• SAP
• SN
Check date:
Indicates when the factory test of
the sensor was carried out.
SAP:
Displays the SAP number of the sensor.
SN:
Displays the serial number of the sensor
electronics.
pH 7.00
Hold
Diag
Select
Error list
Error log
Operation log
Calibration log
Service
Edit [¯]
Next [E]