Appendix
160
a0005016-en
PARAM
Set controller hold
Freeze variables
Yes
(controllers
n/a with hold)
No
Alarm for maximum
dosing time
02 s
(2...9999s)
Selection
On-site calibration
Special buffer
(only pH)
General settings
Calibration timer
Top al calibration
type
Offset
c
Calibration
(only pH)
MTC
ATC
Slope which causes
alarm
(only pH)
25.00mV/pH
(5.00...57.00mV/pH)
Zero point which
causes alarm
1.30 pH/200mV
(0.05...2.00pH /
-500...+500mV))
Cal timer
Cal timer
Warning in
Remaining
time
0001:00
on
0001h
Select hold current
fixed
Min (2,4mA)
Max (22mA)
last
Hold current
(only fixed)
(0...100%)
000%
Set hold after-effect
time
010s
(0...999s)
Assignment to each
error number:
Error no. /
Relay assignment /
Alarm error current /
Cleaning trigger /
Isotherm. comp.
Channel
(only pH)
1
Function 1
off
Function 2
off
Sensor Condition
Check
on
off
Back to return
field
Back to return
field
Selection
Alarm
Hold
Calibration
Validation Topcal
(not available with
Profibus)
Auto hold with:
Cal.
Diag.
Param
CPC
on
on
on
on
Enter alarm delay
0000s
(0...2000 s / min)
Warning, if "Min"
and current range
0...20 mA are
selected
Select
error current
Min (2,4mA)
Off
Max (22mA)
"Return field"
Press the PARAM key
to return to these
highlighted fields
= Code entry required
PARAM
Set up 2
Manual operation
First start up
Set up 1
Special buffer
Edit table for
Buffer no.
1 (1...3)
Enter number of
buffers
2 (2...3)
Special buffer
Enter number of
support points
02 (2...10)
Special buffer: Table
status info field
Invalid table
--> back
valid table
--> back
Back to special
buffer
Edit table... or
se Param to return
to return field
u
Cal parameter
Auto buffer
Fixed buffer
Edit cal. solution
used for calibration
with Topcal
0450mV
(-1500...1500mV)
Buffer type
DIN 19267
....
Special buffer
E+H
Back to return
field
Buffer
Buffer
Buffer
Buffer
1
2.0
4.01
....
6.98
Buffer
Buffer
Buffer
Buffer
2
6.98
9.18
....
4.01
Convey buffer
during calibration
buffer dur. Cal.
Intervall
15s
off
Cal parameters
(only pH)
Data entry
Manual buffer
Automatic buffer
recognition
Fixed buffer
Cal parameters
(only redox abs.)
Calibration abs.
Data entry abs.
Edit Offset
Current PV:
pH/
mV
Offset:
07.00pH/ 0000mV
07.00
0000
Cal parameters
(only redox %)
Data entry rel.
Data entry abs.
Data entry rel.
Data entry abs.
Buffer type
(only pH, fixed
buffer, auto buffer)
DIN 19267
....
Special buffer
E+H
Buffer 1
(only pH)
Buffer
Buffer
Buffer
2.0
4.01
....
6.98
Buffer
(only pH)
Buffer
Buffer
Buffer
2
6.98
9.18
....
4.01
Back to return
field
Back to return
field
Stability calibration
Threshold
02mV
Length
10s
Special buffer
Enter support points
°C
pH
000.00
00.00
Query
Table ok
Delete elements
(then back to the
support points)
Validation alarm
Function
off
Maint.
0.50 pH
on
Validation prog.
Table
Dest.
Val.P2
User prog.1
Val.P1/2 User prog.2
Val+Cal. User prog.3
Val. P1 no prog.