Commissioning
61
7.5.1
Configuring Clean program
The following example illustrates how to configure the Clean cleaning program for your Topcal. All
the cleaning functions and the cleaning and calibration programs are described in the "Set up 2 -
Topcal S" section.
To call the menu, select:
Ñ
> Specialist > Set up 2 > Topcal > Config. Topcal
Exit Quick Setup
Options
• end
• restart
Factory setting
end
Specify whether you want to save the settings
and exit the Quick Setup menu or run through
the menu again to make corrections.
Display
Entry
The current status of the cleaning functions is displayed. Press
F
to continue.
Press
F
to continue.
Select "Prog. editor" and confirm with
F
.
Select "Clean" and confirm with
F
.
Select "Setup" to configure the program steps. Confirm with
F
.
Use the arrows to select the program step you wish to adjust, e.g. Water. Press
F
to edit.
Function
Options
Info
Hold
pH
7. 00
Param
Edit
[¯]
[E]
Next
Start up
end
restart
Clean trigger
off
off
off
Ext. control
pH 7.00
Hold
Param
Status
Next [E]
Automatic
Hold
pH 7.00
Param
Name: V1
Valve 1
Edit [ ¯®]
Next [E]
0...9; A...Z
Set up menu
Cleaning
Pwrfail prg.
Prog. editor
pH 7.00
Hold
Automatic
Edit [¯]
Next [E]
Param
Next [E]
Prog. editor
Clean
Clean C
Clean S
Clean CS
¯ User 1
Param
pH 7.00
Hold
Edit [¯]
Next [E]
Clean
Edit
Insert progr.
Set up
Progr. time
Param
pH 7.00
Hold
Edit [¯]
Change name
EditLine [E]
Clean
3s
120s
60s
60s
pH 7.00
Hold
Param
01 Assembly
service
02 Water
03 Cleaner
04 Wait
05 Water
Select [¯]
¯