Commissioning
92
Aborting programs
A program that has been started (Clean, Clean C, Clean S, Clean CS) goes through the entire cycle
(safety concept). During this time, no other programs can be started.
The service switch on the front door of the control unit has the highest priority. If you move the
switch to the "Service" position, you can also interrupt ongoing programs during operation.
You can interrupt the interval program with a permanent signal at the "Automatic stop" digital input.
The assembly has to be in the "Measure" position for this. The interval program is continued if a
signal is no longer present at the aforementioned input.
Controlling the cleaning and calibration programs via binary contacts
!
Note!
• "1" = Voltage of 10 ... 40 V (duration approx. 400 mS) applied at contacts bin 0 ... bin 2 (terminals
81 ... 86). This auxiliary voltage can be taken from the 15 V auxiliary voltage output of Mycom
S CPM153 for non-Ex devices.
• "0" = 0 V
Program
bin. 0
bin. 1
bin. 2
Term. 81/82
Term. 83/84
Term. 85/86
Clean
(cleaning)
1
0
0
Clean C
(cl calibration)
0
1
0
Clean S
(cl sterilization)
0
0
1
Clean CS
(cl calib sterilization)
1
1
0
User 1
(freely selectable)
1
0
1
User 2
(freely selectable)
0
1
1
User 3
(freely selectable)
1
1
1