Wiring
34
Current output coding
With device versions CPM153-AxA/Bxx (2 current outputs) and CPM153-AxC/Dxx (2 current
outputs with HART), the current outputs can be operated as active or passive outputs. Jumpers on
the controller module M3CH allow recoding.
For non-Ex instruments, these modules may be recoded to active outputs.
a0003983
Fig. 36:
Passive current output 1: Electrical connection of the Commubox FXA191
1
2
3
4
Current output 1
Power unit 24 V DC
Recorder
HART modem Commubox FXA191
5
6
7
PC with operating software
Shielding
Other switching units, PLC with passive input
a0004668
Fig. 37:
Coding for current outputs (Interior view of the
housing upper section)
a0004669
Fig. 38:
Current output coding
A
Coding for passive output
B
Coding for active output
31
33
A
L+
32
34
B
L-
31 33 A L+
32 34 B L-
-
+
³
W
250
1
3
6
7
31
32
5
4
2
1
2
1
1
2
2
1
1
1
1
A
B