Commissioning
99
Select program
Options
• No prog.
• Clean
• Clean C
• Clean CS
• Clean S
• User prog.
Factory setting
Clean
Select the program that should be started if the
network communication or air supply fails.
Activate Topcal
Activate control programs
Activation and subsequent
selection
• Automatic
– On
– Off
• Ext. control
– On
– Off
• Clean trigger
– On
– Off
• Power reset
– On
– Off
Activate the control programs via which a
program should be started.
Activate back pressure water
(sealing water)
Activation and subsequent
selection
• Function
– On
– Off
• Relay
– Water
– Valve 1
– Valve 2
• Length:
01 ... 30 s
Factory setting
Function: On
Relay: Water
Length: 05 s
Specify whether and when sealing water should
be pumped.
Sealing water is pumped into the rinse chamber
before and after moving the assembly. This is
useful in processes with media that are fibrous
or tend to stick as the counterpressure in the
rinse chamber caused by the sealing water
prevents the penetration of medium.
Function:
If you activate the function, sealing
water is pumped into the rinse chamber of the
assembly every time the assembly is moved.
Relay:
Specify via which valve the sealing
water should be pumped in.
• Water: Via the water connection at the
rinsing block
• Valve 1, Valve 2: Via additional valve 1 or 2
Length:
Specify how long sealing water should
be pumped before and after moving the
assembly.
Function
Options
Info
Содержание Topcal S CPC310
Страница 161: ...Appendix 161 ...
Страница 175: ......