Configuring RFC2544 Measurement Settings (AQ1300 Option)
3
2
1
4
5
6
7
8
9
10
11
12
13
14
15
16
App
Index
8.8 Configuring
Back to Back Test Settings
Procedure
Back to Back Test Setup Screen
Follow the procedure in section 8.1 to display the RFC2544 Setup screen.
Press the
Detail Setup
soft key and then the
Back to Back Setup
soft key to display the following
screen.
Set the test duration (1 to 999).
Set the number of trials (1 to 60).
Set the resolution (1 to 65535).
Select this check box to perform
pass/fail judgment.
Set the lower limit on the number of
frames (0 to 1486607142).
Explanation
Test Duration
Set the test duration. The value that you specify here is the time that a single test frame is
transmitted for.
Range: 1 to 999 seconds
Number of Trial
Set the number of trials that you want to perform.
Range: 1 to 60
Resolution(Frames)
Set the range that the result of the back to back test will be contained within.
Range: 1 to 65535 frames
Pass/Fail Judge
Set whether to perform pass/fail judgments on the back to back test result.
• Selected: Pass/fail judgments are performed.
• Cleared: Pass/fail judgments are not performed.
Note
If you want to perform pass/fail judgments, enable the pass/fail judgment setting in the optional measurement
settings as well.
Threshold Criterion
If you select to perform pass/fail judgments on the back to back test result, set the judgment
threshold.
Range: 0 to 1486607143 frames