8.7 Configuring
Frame Loss Rate Test Settings
Procedure
Frame Loss Rate Test Setup Screen
Follow the procedure in section 8.1 to display the RFC2544 Setup screen.
Press the
Detail Setup
soft key and then the
Loss Rate Setup
soft key to display the following
screen.
Set the test duration (1 to 999).
Set the number of trials (1 to 60).
Select this check box to have the rate
decreased between measurements.
Set the initial rate (1 to 100).
Set the step down rate (10%, 20%).
Select this check box to perform
pass/fail judgment.
Set the upper limit on the loss rate
(0.01 to 100.00).
Explanation
Test Duration
Set the test duration. The value that you specify here is the time that a single test frame is
transmitted for.
Range: 1 to 999 seconds
Number of Trial
Set the number of trials that you want to perform.
Range: 1 to 60
Initial Rate
Set the initial traffic rate that the frame loss rate test will use.
Range: 1 to 100%
Step Down
Set whether the rate is decreased between measurements in the frame loss rate test.
• Selected: The rate is decreased between measurements.
• Cleared: The rate is not decreased between measurements.
Note
The test result format is different depending on whether the rate is decreased between measurements.
For details, see section 9.5, “Displaying Frame Loss Rate Test Results.”