8-3
IM AQ1300-01EN
Configuring RFC2544 Measurement Settings (AQ1300 Option)
3
2
1
4
5
6
7
8
9
10
11
12
13
14
15
16
App
Index
Explanation
Latest Setup
Select this item to use the setup that was displayed previously.
Default Setup
Select this item to return to the default setup.
File
Select this item to load a setup file (with an .sd extension) from the file list.
Select this item when you want to load a setup file that is not in the setup file list.
To create a setup file, you can use the setup software and send the file to the AQ1300/AQ1301, or
you can save the settings on the AQ1300/AQ1301.
Setup File Selection
To select a registered setup file, use the setup file list.
You can register up to 48 setup files to the setup file list (4 pages with 12 files per page).
A comment or file name is displayed for each of the registered setup files in the setup file list.
The setup file list is updated when you open the Select Setup File screen after you have created a
display management file and setup files with the setup software and sent the files to the AQ1300/
AQ1301.
For details, see the
Setup Software User’s Manual
, IM AQ1300-61EN.
Note
• The AQ1300/AQ1301 can only display the setup file list if the display management file that you created
using the setup software (disManage.dmf) and the setup (.sd) files that the display management file refers
to are saved to the AQ1300/AQ1301 /setup directory.
Example: /setup/disManage.dmf
/setup/0000.sd
/setup/0001.sd
:
• When you are performing RFC2544 tests, you can only select setup files for RFC2544 tests.
8.1 Selecting a Setup File