8-26
IM AQ1300-01EN
Operation after Measurement Stops
Judge Pass or fail
Select whether to perform pass/fail judgments after measurements finish.
• Selected: Pass/fail judgments are performed after measurements finish.
• Cleared: Pass/fail judgments are not performed after measurements finish.
Note
To perform pass/fail judgments for a particular test, you have to set the pass/fail judgment setting on the
test’s setup screen.
Save measurement results
Select whether to save the measurement results to a file after measurement finishes.
• Selected: The measurement results are saved to a file when measurement finishes.
• Cleared: The measurement results are not saved to a file when measurement finishes.
Save measurement logs
Select whether to save the measurement log to a file after measurement finishes.
• Selected: The measurement log is saved to a file when measurement finishes.
• Cleared: The measurement log is not saved to a file when measurement finishes.
Others
Subtrac the fixed latency from the result
During latency measurement, the fixed delay that occurs in the other device during loopback is
subtracted from the measured results. The subtracted results are displayed as measurement
results. If the measured value is less than the fixed delay, 0.00 μs is displayed.
This feature is supported in firmware version (FW Ver.) R1.08.01.001 and later.
• Selected: The fixed delay is subtracted from the measured values.
• Cleared: The fixed delay is not subtracted from the measured values.
Fixed delay value based on the interface
Interface
Delay
XFP
1.0 µs
SFP (GbE)
1.4 µs
SFP (FE)
12 µs
RJ-45 (1000M)
1.6 µs
RJ-45 (100M)
11 µs
RJ-45 (10M)
108 µs