4-28
IM AQ1300-01EN
4.11 Configuring Options (Auto)
Procedure
Advance Setup Screen
Follow the procedure in section 4.1 to display the Auto Setup screen.
Press the
Option (Auto)
soft key and then the
Advance setup
soft key to display the following screen.
Set the starting test number (1-8).
Set the interval between the pressing of
START and the start of transmission
(1 to 10 s).
This setting is valid when the “Synchronize
Measurement and Transmission controls”
check box is selected.
Select whether or not to check before
proceeding to the next test item
whenever a test item has finished (ON,
OFF).
Set the interval between test items
(1-10 s).
These settings are valid when
Execution Type is set to Continue.
Set the test item execution type (Single,
Continue).
Method for saving continuance test
result files (Separate, Consolidate)
Measurement Screen
Press the
Option (Auto)
soft key and then the
Measurement Setting
soft key to display the following
screen.