6-54
IM AQ1300-01EN
When L3 Error is detected
Select whether or not to stop measurement when an L3 error is detected during measurement.
• Selected: Measurement is stopped when an L3 error is detected.
• Cleared: Measurement is not stopped even when an L3 error is detected.
Operation after Measurement Stops
Save measurement results and Statistics Log
Select whether or not to save the measurement results and the statistics log to a file after
measurement finishes. The statistics log is only saved to a file when the “Execute Logging” check
box is selected.
• Selected: The measurement results and the statistics log are saved to a file when measurement
finishes.
• Cleared: The measurement results and the statistics log are not saved to a file when
measurement finishes.
Note
The measurement results and the statistics log are saved to the directory that you specify in the File Name
Setup in the Measurement menu with the same file name but different extensions.
Example
When the file name is test_0001
Statistical result file: test_0001.mr
Statistical log file: test_0001.csv
Synchronization Control Setting
Try synchronizing with the other device
Select whether or not to try synchronizing with the other device (the address specified as the
destination address) when the START or STOP key is pressed. When the START or STOP key is
pressed, the AQ1300/AQ1301 sends a measurement start or stop packet to the other device at the
destination address.
• Selected: The AQ1300/AQ1301 tries to synchronize with the other device.
• Cleared: The AQ1300/AQ1301 does not try to synchronize with the other device.
Note
When you are performing BERT test, the AQ1300/AQ1301 does not send measurement stop packets.
Others
Subtrac the fixed latency from the result
During latency measurement, the fixed delay that occurs in the other device during loopback is
subtracted from the measured results. The subtracted results are displayed as measurement
results. If the measured value is less than the fixed delay, 0.00 μs is displayed.
This feature is supported in firmware version (FW Ver.) R1.08.01.001 and later.
• Selected: The fixed delay is subtracted from the measured values.
• Cleared: The fixed delay is not subtracted from the measured values.
Fixed delay value based on the interface
Interface
Delay
XFP
1.0 µs
SFP (GbE)
1.4 µs
SFP (FE)
12 µs
RJ-45 (1000M)
1.6 µs
RJ-45 (100M)
11 µs
RJ-45 (10M)
108 µs