Error
Number
Message
Solution
1280
Inband Session is disconnected.
Check the network settings of the other device.
1281
Incorrect date and time setting.
Set the correct date and time.
Set the correct date and time.
1282
Incorrect User Name setting.
Set the correct User Name.
Set the user name using up to 15 characters.
1284
The control to the destination machine
failed.
Check the source/destination address and the network settings of
the other device. If you do not want to use the function that tries to
synchronize to the other device, disable this function in the test option
settings.
1285
The control to the destination machine
failed.
There is a possibility that is not the
measurement result display.
Press STOP on the other measuring device to display the measurement
screen, and then start measurement again. If you do not want to use the
function that tries to synchronize to the other device, disable this function
in the test option settings.
1286
It failed in making of the file for the screen
forwarding.
Wait a moment, and then try again.
1287
It failed in forwarding the screen.
Check the source/destination address and the network settings of the
other device.
1288
Cannot make file or folder in this folder.
Delete unnecessary files and directories.
1289
Cannot make statistics log file, because log
data doesn’t exist.
Measure first and then save. Or, check the statistics log settings.
1290
The system version of the destination is
mismatch.
Check the version of the other device.
1291
Cannot make measurement result file,
because measurement data doesn’t exist.
Measure first and then save.
1292
Load Error(Bad Measurement Type).
(This File is for “RFC2544”.)
Match the test type that you are loading to the test type that you used to
create the test and then load.
1293
Some sd files were not found.
Check that all the setup files (sd files) that are defined in the setup file list
are in the specified directory.
1294
Decompression of the dmfz file failure.
Create the dmfz file again using the setup software.
1295
This File is for “10G”. The test interface
was changed into SFP from XFP.
To test 10G, use the AQ1300.
1296
Load Error(Bad Measurement Type)
(This File is for “VLAN Test”.)
Match the test type that you are loading to the test type that you used to
create the test and then load.
1297
The file format is not valid.
Check the contents of the file.
5001
System Error.
The system may be broken.
Contact your nearest YOKOGAWA dealer.
5002
FPGA Configuration Error.
5003
FAN Alarm.
5004
Detect Bad Clock.
5005
Detect Bad Optical Module.
The optical module may be broken.
Replace it with a new optical module.
16.2 Messages and Corrective Actions