Appendix B: Performance Verification
DG2030 User Manual
BĆ11
Setup.
H
Oscilloscope
Displayed channel
CH1
Vertical axis
500 mV/div
. . . . .
Horizontal axis
1
m
s/div
. . .
Trigger mode
Auto
. . . .
Trigger level
500 mV
. . . . .
Input coupling
DC
. . . .
Input impedance
50
W
. .
H
Function generator
Waveform
Square wave
. . . . . . .
Frequency
1 MHz
. . . . . . .
Amplitude
1 V (50
W
termination)
. . . . . . .
Offset
500 mV (50
W
termination)
. . . . . . . . . .
Characteristics Confirmation Procedure.
1. Load the TP2EXCLK.PDA test pattern file from the performance check
disk.
2. Press the START/STOP button on the front panel so that the button’s LED
indicator lights.
3. Confirm that the clock pulse waveform can be observed on the oscilloscope.
4. Press the START/STOP button on the front panel to stop output (the
button’s LED goes off).