Utility Menu
3Ć100
DG2030 User Manual
Table 3-22 lists the error codes, their meaning, and failure board for the error
code. See the DG2030 Service Manual for detailed descriptions of the error
codes.
Table 3Ć22: Error Code
Error code
Description
Failed component
1XX
CPU diagnostics error
A6 CPU board
2XX
Display diagnostics error
A6 CPU board
3XX
Front panel diagnostics error
A12 Key board
4XX
Clock diagnostics / calibration error
A30 Clock board
5XX
Trigger diagnostics / calibration error
A30 Clock board
6XX
Sequence memory error
A50 PG&Output board
7XX
Pattern memory diagnostics error
A50 PG&Output board
Side Menu
Item
Function
Type
Selectsthe diagnosticstest. Either individual itemsor all tests(All) can
be selected. An asterisk is displayed next to selected items.
Execute
Executes the diagnostic tests for the items marked with an asterisk.
Operation.
Diagnostics
Bottom button
Popup menu
Side button
Diag
Type (Select a diagĆ
nostic test or All with
the general purpose
knob.)
Execute