Appendix B: Performance Verification
DG2030 User Manual
BĆ17
The following steps check the digital pattern output, event output and sync
output in Enhanced run mode.
17. Press the following buttons to set the DG2030 run mode to Enhanced.
Menu button
Bottom button
Pop up menu
Side button
Front panel
button
SETUP
Run Mode
Enhanced
18. Change the oscilloscope trigger source to CH3 and trigger position to 50 %.
19. Press the following buttons to set the trigger parameters.
Menu button
Bottom button
Pop up menu
Side button
Front panel
button
SETUP
Trigger
Source
(Set to Int)
Interval
State (Set to On)
Time
10
kHz/ms/mV
20. Press the START/STOP button on the front panel so that the button’s LED
indicator lights.
21. Confirm the following points from the oscilloscope screen:
H
The output patterns in the CH0 to CH3 correctly appear as shown in
Figure B-8.
Change the CH0 cable connection to, in turn, from CH1 to CH3 cable
connection when you observe the outputs from CH1 to CH3.
H
The output patterns in the CH4 to CH7 correctly appear as shown in
Figure B-8 if the option 01 has been installed.
Change the CH3 cable connection to, in turn, from CH4 to CH7 cable
connection on the rear panel when you observe the outputs from CH4 to
CH7.