Operating Examples
DG2030 User Manual
2Ć45
4. Set the sampling clock frequency to 50 MHz.
Menu button
Bottom button
Popup menu
Side button
Front panel
button
Clock
Source
(Select Int.)
Int FREQ
50
MHz
PLL
(Select On.)
5. Set the signal generation mode to continuous mode.
Menu button
Bottom button
Popup menu
Side button
Front panel
button
Run Mode
Repeat
6. Set the channel CH0 to CH7 output levels to 3 V for the high level and –1 V
for the low level.
Menu button
Bottom button
Popup menu
Side button
Front panel
button
Level Condition
Select channel
CH0
by pressing the front panel up and down arrow buttons.
High
3
ENTER
Low
-1
ENTER
Set the output levelsfor channels
CH1
to
CH7
in the same manner.
Setting Sampling Clock
Frequency
Setting Signal Generation
Mode
Setting Output Level