Appendix B: Performance Verification
DG2030 User Manual
BĆ5
Self Tests
Execute the DG2030 self tests and confirm that no errors occurred.
1. Press the required buttons in the following order. Selecting All from the
Type side button causes the instrument to run the self-test on all test items.
Menu button
Bottom button
Pop up menu
Side button
Front panel
button
UTILITY
Diag
Type (Select All)
Execute
2. Confirm that all self tests passed by checking the Status display area shown
in Figure B-2.
Figure BĆ2: Diagnostics menu
If Fail is listed in the Status display area, an error code will be displayed in
the Comment column. See Table B-3 for the meanings of the error codes.
NOTE
. Contact your Tektronix sales representative if an error occurs.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com