Getting Started
1Ć10
DG2030 User Manual
ON/STBY Switch
Figure 1Ć2: Location of the ON/STBY switch
StartĆup Diagnostics
8. The DG2030 automatically runs diagnostics when the instrument is turned
on from the ON/STBY switch. These diagnostics check whether the
instrument is performing within its defined operating characteristics. If all
the diagnostic items have been completed without error, the instrument
displays the EDIT menu.
If an error is detected, the instrument displays a fail and error code message.
You can exit this state by pressing any front-panel button to display the
EDIT menu. However, until the error is corrected, the instrument perfor-
mance cannot be relied on.
If the instrument chassis temperature is outside the specified operating range,
an error will occur during the power-up diagnostics. If this happens, turn off
the instrument, wait until the chassis temperature is within normal operating
range, and then switch the power on again.
NOTE
. Contact your local Tektronix Field Office or representative if an error is
displayed.