Appendix B: Performance Verification
DG2030 User Manual
BĆ29
Setup.
H
Digital multi–meter
Function
DCV
. . . . . . . . . . . . . . . . .
Range
Auto
. . . . . . . . . . . . . . . . . . .
H
Oscilloscope
Vertical axis
500 mV/div
. . . . . . . . . . . . . .
Horizontal axis
2 ns/div
. . . . . . . . . . . .
Trigger source
CH1
. . . . . . . . . . . . .
Trigger level
750 mV
. . . . . . . . . . . . . .
Display channel
CH1
. . . . . . . . . . .
Input coupling
DC
. . . . . . . . . . . . .
Input impedance
50
W
. . . . . . . . . . .
Bandwidth
Full
. . . . . . . . . . . . . . .
High Ref. (Rise/Fall Times)
80 %
. .
Low Ref (Rise/Fall Times)
20 %
. . .
Characteristics Confirmation Procedure.
1. Load the TP7LVL.PDA test pattern file from the performance check disk.
The following steps check the high voltage levels on all output channels.
2. Press the following buttons to set the DG2030 high and low voltage levels.
Menu button
Bottom button
Pop up menu
Side button
Front panel
button
SETUP
Level Condition
Move the cursor to CH0 with the general purpose knob.
High
-1.25
ENTER
Low
-1.50
ENTER
Note that the low level is automatically set to –1.50 V when the high level
has been set to –1.25 V.
3. Press the START/STOP button on the front panel so that the button’s LED
indicator lights.