Appendix B: Performance Verification
DG2030 User Manual
BĆ27
10. Set the inhibit control for the CH2 output channel to OFF as following
procedures:
Menu button
Bottom button
Pop up menu
Side button
Front panel
button
SETUP
Select CH2 with the general purpose knob.
Output
Condition
Control
Condition
Change Inhibit
Control
OFF
OK
11. Confirm that the clock output voltage level is almost 0 V.
12. Disconnect the BNC cable from the inhibit input on the DG2030 rear panel.
13. Confirm that the pulses with 10 MHz of frequency, +3 V of high voltage
level and –1 V of low voltage level are continuously output from the
DG2030 clock output.
14. Press the START/STOP button on the front panel to stop output.