Appendix B: Performance Verification
DG2030 User Manual
BĆ15
The following steps check the pattern data output operation in Step run mode.
12. Press the following buttons to set the DG2030 run mode to Step.
Menu button
Bottom button
Pop up menu
Side button
Front panel
button
SETUP
Run Mode
Step
13. Cancel the single acquisition mode setting in the oscilloscope, and then start
acquisition.
This can be made by selecting SHIFT + ACQUIRE (front panel
!
Stop
After (bottom)
!
Run/Stop button only (side) for the TDS784A, and then
by pressing the RUN/STOP button.
14. Press the START/STOP button on the front panel so that the button’s LED
indicator lights.
15. One bit appears in the CH0 output each time the STEP/EVENT button on
the DG2030 front panel is pressed. Press the STEP/EVENT button six
times and confirm that the bit pattern: 0 1 0 1 0 1, one bit for every pressing
that button, appears.
16. Press the START/STOP button on the front panel to stop output (the
button’s LED goes off).