Appendix B: Performance Verification
BĆ12
DG2030 User Manual
This test confirms that pattern data is output from the DG2030 output channel
ports (CH0 through CH7) in various run modes.
Equipment Required
Oscilloscope
Function generator
Five BNC cables
BNC T adapter
Performance check disk
Connections.
Connect the clock output from the DG2030 front panel to the
oscilloscope CH1 input with an BNC cable. Connect the function generator
output to the DG2030 rear panel event input with the other BNC cable.
Next, using the BNC cables, connect the CH0 signal output from the DG2030 to
the oscilloscope CH2 input, the event output from the DG2030 rear panel to the
oscilloscope CH3 input, and the sync output from the DG2030 rear panel to the
oscilloscope CH4 input. See Figure B-5 for those connection
DG2030 Front Panel
Clock Output
BNC Cable
TDS784A Front Panel
CH1
AFG2020 Front Panel
BNC Cable: connect to the
AUX INPUT:
EVENT
connector on the DG2030 rear panel.
BNC Cable: connect to the
AUX OUTPUT:
EVENT
connector on the DG2030 rear panel.
BNC Cable: connect to the
AUX OUTPUT:
SYNC
connector on the DG2030 rear panel.
CH2
CH3
CH4
CH0
Figure BĆ5: Sequence & digital output connections #1
Sequence & Digital Output
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com